C-4
Alarm
Alarm
Reason Countermeasure
acknowledge the alarm
condition.
[Time Near End] Low
Infusion remaining time
reaches the setting value of
[Time Near End].
1. The alarm will not be
cancelled automatically
until the infusion is
completed, and then
switch to [VTBI Complete]
or [Empty] alarm.
2. Or press to cancel
alarm.
3. Or press to
acknowledge the alarm
[No AC Power] Low
Power cord disconnects when
the pump has connected to an
AC/DC power supply.
1. Connect to an AC/DC
power supply to cancel the
alarm.
2. Or press to cancel
[Communication
interrupted]
Low
Communication is abnormally
interrupted for 3 minutes (due
to pump been pulled out or
internal abnormality) after
pump inserted to BeneFusion
DS5 Infusion Supervision
System and communicated
successfully with the system
(no pump serial number
conflict or version number
compatible), or the pump and
CIMS are communicated
successfully over Wi-Fi
, the
network communication is
abnormally interrupted for 3
minutes. After the alarm is
triggered, infusion of the pump
will not be influenced, and the
Press or insert the
pump to BeneFusion DS5
Infusion Supervision
System, or restore the
communication between
the pump and CIMS.