6-10 BeneVision N22/BeneVision N19 Patient Monitor Service Manual
6.4.6 Removing LED Board/Indicator Board
As shown in the following picture, loosen and remove the two M3X6 cross recessed pan head screws with pad, then
remove the indicator board.
As shown in the following picture, loosen and remove the four M3X6 cross recessed pan head screws with pad, then
remove the LED board.
6.5 Disassembling Main Unit
6.5.1 Removing iView Assembly (iView Assembly Optional)
As shown in the following picture, loosen and remove the two M3X6 cross recessed pan head screws with pad, then
remove the iView assembly.
M3 screw
M3screw
M3 screw