4 Units Description
4-
45
3
Check whether the five electrodes are installed properly.Make sure the
electrodes stand in a line vertically and their front surfaces are lined up.
Remove the electrodes following the reverse steps described above. It is important to
execute the procedure such as tubing purge and warm-up before dismounting the
electrodes. Store the electrodes properly after removing them from the instrument. Refer to
Chapter 7.7.4 for more details about removing and storing the electrodes.
CAUTION
The ISE unit (optional) should be on power all the time. In some cases the
Power will be shut down for more than half an hour, strore the electrodes
referring to Chapter 7.7.4.