4 Units Description
4-44
4.8.3.4 ISE Electrodes
The steps of installing the ISE Electrodes are as follows.
1 Take out the reference electrode from the package and pull out the yellow
intubatton. Pull down the press plate of the ISE module and push the reference
electrode into the ISE module. (Don’t throw away the yellow intubatton and
use it again when storing the electrode.) If white crystalloids adhere to the tube
of the reference electrode, wash it with warm water before installing.
2
Put the other electrodes into the ISE module. It is easier to install the K+, Na+,
Cl- electrodes than the spacer electrode. To insert the spacer electrode (the
top one), it is needed to slightly press the press board.
Caution: Replace the space electrode with the Li+ electrode when necessary.