9 Troubleshooting                                                                                    9-9 
400000730000
Operating  software 
error:  Receiving  buffer 
overflows 
Receiving buffer overflows  Please contact the R&D engineer. 
400000810001
Test  result  error:  No 
balance  point  found 
in … 
Reagent,  sample  error  or 
test parameter configuration 
error. 
Rerun the test after replacing the reagent or sample. If the 
test  parameters  are  wrong,  please  reconfigure 
them.Check whether the optical module is wrong. 
400000820001
Test  result  error:  No 
linear range found in … 
Reagent  not  stable  or 
something  wrong  with  the 
sample 
Rerun the test after replacing the reagent or sample.If the 
test  parameters  are  wrong,  please  reconfigure 
them.Check whether the optical module is wrong. 
400000830001
Test  result  error: 
Linearity  of  reaction 
curve of … too weak 
Reagent  not  stable  or 
something  wrong  with  the 
sample 
Rerun  the  test,  or  rerun  after  the  reagent  or  sample  is 
replaced.If  the  test  parameters  are  wrong,  please 
reconfigure  them.  Check  whether  the  optical  module  is 
wrong. 
400000840001
Test  result  error: 
Response  of  … 
calculation error 
Data of critical points lost. 
Rerun  the  test.  Check  whether  the  optical  module  is 
wrong. 
400000850001
Test  result  error: 
Response of … exceeds 
the  one  of  weakest 
calibrator 
Sample  abnormal 
(hemolysis);  calibrator 
concentration too high 
Make sure the calibrator and calibration rule configuration 
are  ok.  If  not,  change  the  configuration  and  rerun  the 
calibration and QC. If the sample is abnormal, rerun after 
the  sample  is  processed.  Check  whether  the  optical 
module is wrong. 
400000860001
Test  result  error: 
Response of … exceeds 
the  one  of  strongest 
calibrator 
Sample  abnormal 
(hemolysis);  calibrator 
concentration too low 
Make sure the calibrator and calibration rule configuration 
are  ok.  If  not,  change  the  configuration  and  rerun  the 
calibration and QC. If the sample is abnormal, rerun after 
the  sample  is  processed.  Check  whether  the  optical 
module is wrong. 
400000870001
Test  result  error: 
Concentration  of  … 
exceeds the low limit of 
linear range 
Reagent  not  stable  or 
something  wrong  with  the 
sample 
Rerun  the  test,  or  rerun  after  the  reagent  or  sample  is 
replaced.If  the  test  parameters  are  wrong,  please 
reconfigure  them.  Check  whether  the  optical  module  is 
wrong. 
400000880001
Test  result  error: 
Concentration  of  … 
exceeds the high limit of 
linear range 
Reagent  not  stable  or 
something  wrong  with  the 
sample 
Rerun  the  test,  or  rerun  after  the  reagent  or  sample  is 
replaced.  If  the  test  parameters  are  wrong,  please 
reconfigure  them.  Check  whether  the  optical  module  is 
wrong.