9 Troubleshooting 9-11
400001020001
Test result error:
Calibration SD of too
large
Calibration SD too high
Replace the reagent and calibrators. Confirm the
calibrators and calibration rule. Recalibrate. If the error
remains, check the optical system.
400001030001
Test result error:
Difference between
calibration coefficients
of … too large
Something wrong with the
calibrator and reagent
Replace the reagent and calibrators. Confirm the
calibrators and calibration rule. Recalibrate. If the error
remains, check the optical system.
400001040001
Test result error:
Calibration related
coefficients of … too low
Something wrong with the
calibrator and reagent
Replace the reagent and calibrators. Confirm the
calibrators and calibration rule. Recalibrate. If the error
remains, check the optical system.
400001050001
Test result error:
Incomplete repeated
calibration data of …
Something wrong with the
calibrator and reagent
Replace the reagent and calibrators. Confirm the
calibrators and calibration rule. Recalibrate. If the error
remains, check the optical system.
400001060001
Test result error:
Calibration curve of
not monotonic
Something wrong with the
calibrator and reagent
Replace the reagent and calibrators. Confirm the
calibrators and calibration rule. Recalibrate. If the error
remains, check the optical system.
400001070001
Test result error:
Concentration of
calculation failed
Something wrong with the
calibrator and reagent
Replace the reagent and calibrators. Confirm the
calibrators and calibration rule. Recalibrate. If the error
remains, check the optical system.
400001080001
Test result error:
Incomplete test result
of …
Rerun calibration not
completed. or insufficient
reagent and calibrator.
Refill the reagent and calibrator; recalibrate.
400001090001
Test result error: Error of
repeated … too large
Something wrong with the
calibrator and reagent
Replace the reagent and calibrators. Confirm the
calibrators and calibration rule. Recalibrate. If the error
remains, check the optical system.
400001100000
Test result error: QC out
of control
Control performance
degraded; something wrong
with the reagent
Replace the reagent and calibrators. Confirm the
calibrators and calibration rule.Rerun.If the error remains,
check the optical system.
400001140006
Operating software
error: Detergent
exhausted, or invalid
sensor
Wash solution used up or
sensor invalidated
Check the DI tank. If there is insufficient DI water, refill it
and execute failure recovery.
400001150006
Operating software
error: Waste full
Waste tank full
Check the waste tank. If it is full, empty it and execute
failure recovery.
400001350000
Test result error: Blank
response of … too low
Blank reponse too low Check the cuvette, reagent and diluent.