EasyManua.ls Logo

Mindray BS-120 - Page 213

Mindray BS-120
267 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 Troubleshooting
9-20
100650030005
Reaction unit result
error: Mechanical
resetting error
Mechanical resetting error
Check the motor/sensor/cable. Run mechanical resetting.
If the error remains, restart the analyzing unit and the
operating software. Rerun mechanical resetting again.
100650040005
Reaction unit result
error: Invalid status.
Self-check
Self-checking; Other
operations not supported
Execute the previous operation after waiting for 30-60
seconds. If there is no response for a long time, execute
mechanical resetting or restart the analyzing unit and the
operating software.
100650040015
Reaction unit result
error: Invalid status.
Error
Reaction unit error; current
operations not supported
Execute failure recovery. Check the motor/sensor/cable. If
the error remains, please contact the R&D engineer.
100650040025
Reaction unit result
error: Invalid status.
Waiting for handshake
Unit busy. No response
Execute the previous operation after waiting for 30-60
seconds. If there is no response for a long time, execute
mechanical resetting or restart the analyzing unit. Rerun
the operating software.
100650050005
Reaction unit result
error: Unit busy. No
response
Unit busy. No response
Execute the previous operation after waiting for 30-60
seconds. If there is no response for a long time, execute
mechanical resetting or restart the analyzing unit and the
operating software.
100650090005
Reaction unit result
error: Wrong
configuration parameter
Parameter configuration
error
Configure the right parameter.
100650110005
Reaction unit result
error: Undefined search
Undefined search Input the right searching parameter.
100650130005
Reaction unit result
error: Wrong system
operation parameter
Wrong system operation
parameter
Input the right operation parameter.
100650140005
Reaction unit result
error: Configure
undefined parameter
Reaction unit error. Can
not find home position
Check the motor/sensor/cable. Check the movement of
the reaction disk by using Aligment. If the error remains,
contact the R&D engineer.