9 Troubleshooting 9-21
100650140015
Reaction unit result
error: Rotation error.
Cannot move away from
the home position
Reaction unit error. Can
not find home position
Check the motor/sensor/cable. Check the movement of
the reaction disk by using Aligment. If the error remains,
contact the R&D engineer.
100650140025
Reaction unit result
error: Rotation error.
Step missing
Reaction unit error. Can
not find home position
Check the motor/sensor/cable. Check the movement of
the reaction disk by using Aligment. If the error remains,
contact the R&D engineer.
100650140045
Reaction unit result
error: Rotation error.
Mixing
Mixing. Rotation not
permitted
Rotate the reaction disk after mixing is finished and the
mixing bar leaves the cuvettes.
100650140055
Reaction unit result
error: Rotation error.
Aspirating/dispensing
sample
Aspirating/dispensing
sample. Rotation not
permitted
Rotate the reaction disk after sample
aspirating/dispensing is finished.
100650140065
Reaction unit result
error: Rotation error.
Aspirating/dispensing
reagent
Aspirating/dispensing
reagent. Rotation not
permitted
Rotate the reaction disk after reagent
aspirating/dispensing is finished.
100650140075
Reaction unit result
error: Rotation error.
Aspirating/dispensing
R2
Aspirating/dispensing R2.
Rotation not permitted
Rotate the reaction disk after R2 aspirating/dispensing is
finished.
100650150005
Reaction unit result
error: Photoelectric
error. Lamp off
Photoelectric error. Lamp
off
After adjusting the lamp by using Alignment, run
photoelectric collection. If the error remains, please
contact the R&D engineer.
100650150015
Reaction unit result
error: Photoelectric
error. Lamp too dark
Lamp intensity not enough
or lamp off
Check the status of the lamp and its connection. Check
the installation of the lamp. Replace the lamp if necessary.