7 - 36 Operator’s Manual
7 3D/4D
• The automatic measurement results appear only for these having the same
characteristics on the planes.
• Once the automatic measurements are completed, the operations to axial rotation,
reference line rotation, parameter adjustment, MSP editing, zooming/panning, dual/
quad-split display will remove the measurement results.
a. Tap [Edit] to modify the measurements. The caliper becomes green.
Or, press <Set> to activate the caliper (becoming green).
b. Move the trackball and press <Update> to modify the length and the position of the
caliper.
c. Press <Set> to confirm the caliper. The caliper becomes white. [Edit] is off.
9. Tap [Auto Comment], the system adds the orientation and the organ comments to the desired
area according to the active ultrasound image.
NOTE:
• The orientation comments describe the location of the plane, referring to A (anterior), P
(posterior), L (Left), R (right), U (up), D (down).
• Organ comments describe the position of the organ, referring to CSP (cavum septum
pellucidum), T (thalamus), CH (cerebellar hemisphere), CV (cerebellar vermis), CM
(cisterna magna), LV (lateral ventricles).
a. Tap [Font Size] to adjust the font size of the comment.
b. see “14.2 Comments” for adding, moving or deleting the comments.
c. Save the image.
d. Tap [Auto Comment] again to clean them.
10. Tap [Accept Result] to save the measurements to the report.
11. Add the comment and body mark on the plane. Perform the measurement, and save the single
frame/multi-frame image.
7.15.2 Smart ICV
Smart ICV is used to measure fetal cerebral volume.