14 Measurement, Comments and Body Mark
Operator’s Manual 14 - 3
Within the selected field range, the measurement accuracy is ensured within the range mentioned
above. The accuracy specifications are performance in the worst conditions, or based on the real
test for the system, regardless of acoustic speed error.
14.2 Comments
Comments can be added to an ultrasound image to bring attention, annotate or communicate
information observed during the examination. You can add comments to: zoomed image, cine
review image, real-time image, frozen image. You can type the character as comments; insert the
pre-defined comments from the comment library; insert arrow markers or add the trace.
You must ensure that the entered comments are correct. Incorrect comments
may lead to misdiagnosis.
14.2.1 Touch Screen Displaying
• Set the start point of the comment cursor.
Move the cursor to the desired position, tap [Set Home].
• Return the cursor to the set home location.
Tap [Home], the cursor returns to the start position.
• Navigate through comments libraries
Tap [Library] to select the comment library.
• Set comments
Tap to enter
Smart Hip Within ±20%
Smart HRI Within ±5%
Smart B-line Within ±20%
AutoEF Within ±10%
RIMT Within ±10%
IMT Within ±10%
Smart Trace Within ±20%
Smart Fetal HR Within ±10%
Smart Calc Within ±10%
Smart Planes CNS Within ±20%
Smart Pelvic Within ±20%
Smart Bladder Within ±10%
R-VQS Within ±10%
Measurement Item Error