17 - 34 Operator’s Manual
17 Probes and Biopsy
NGB-018
Perform the following procedure:
1. Install the needle-guided bracket:
a. Put on the sterile probe sheath.
b. Select a suitable needle-guided bracket and match the groove to the tab of the probe.
Mount the bracket onto the probe.
c. Screw the pinch nut of the needle-guided bracket to ensure that the needle-guided bracket
is properly installed on the probe.
2. Adjust the needle angle to the proper shift as required:
a. Loosen the angle pinch nut.
b. Adjust the angle block to the desired level.
c. Tighten the angle pinch nut.
3. Install the guiding block:
1 Pinch nut of bracket 2 Angle pinch nut
3 Angle block 4 Clamp
5 Groove 6 Angle-adjusting base
7 Angle shift sign 8 Guiding block specification
9 Guiding block 10 Needle guide hole
1
2
3
4
5
6
7
8
9
10