Information Codes and Error Codes Operation
2 - 16 0070-10-0603-01 Duo™ Operating Instructions
2.4.2 Error Codes
MESSAGE TYPE CODE DESCRIPTION REASON
NIBP E01 NIBP SELF TEST ERR NIBP module hardware failure.
E02 NIBP COMM ERR Communication with NIBP module has
failed.
E04 AIR LEAK Cuff, hose or connector is damaged.
Internal leak.
E05 AIR PRESSURE FAILURE Stable pressure value is not available.
(e.g., hoses are pinched or occluded)
E12 NIBP SYSTEM FAILURE Operation of blood pressure pump system
failed.
E14 *NIBP ILLEGALLY
RESET
Unexpected NIBP reset.
E15 NIBP RESET FAILED NIBP reset failed.
E16 *NIBP COMM CRC
ERROR
NIBP Serial Communication failure
E17 NIBP PATIENT SIZE
CHANGE ERR
Attempt to change patient size failed
MASIMO SPO2 E24 SPO2 COMM ERROR The monitor and the SpO2 modules are
not communicating properly.
E25 SPO2 BOARD FAULT Masimo SET board failed to operate
properly.
E26 SPO2 SENSOR FAULT Defective sensor.
NELLCOR SPO2 E42 SPO2 COMM ERROR The front end module is having problems
communicating (i.e., framing errors or
bad checksums) with the Nellcor board.
E45 SPO2 BOARD FAULT The SpO2 board malfunctions.
GENERAL/
TECHNICAL
E504 KEYBOARD ERR1 Error with front panel keypad board.
E505 MONITOR SHUTOFF
FAILURE
Monitor cannot be turned off normally
E506 SPO2 MODULE NOT
RECOGNIZED
Monitor cannot communicate with the
SpO2 module during self-test.