Patient Parameter Specifications Appendix
5 - 4 0070-10-0603-01 Duo™ Operating Instructions
5.2 Patient Parameter Specifications
5.2.1 NIBP Sub-System Performance Characteristics
The NIBP function is capable of providing non-invasive systolic, diastolic and mean blood
pressure measurements in Pediatric and Adult modes using a blood pressure cuff.
The NIBP function is in accordance with the requirements of EN 1060-1, EN 1060-3 and
ANSI/AAMI SP-10:1992.
5.2.2 Systolic Pressure Measurement
5.2.3 Diastolic Pressure Measurement
* Blood pressure measurements determined with this device are equivalent to those
obtained by a trained observer using the cuff/stethoscope auscultation method, within
the limits prescribed by ANSI/AAMI SP-10:1992, Electronic or automated
sphygmomanometers.
NOTE: Mean Arterial Pressure (MAP) is defined as:
Mean Pressure 1 = Mean Pressure determined from the
oscillometric profile
Mean Pressure 2 = (2*diastolic + systolic) / 3
Mean Pressure Displayed = (Mean Pressure 1 + Mean
Pressure 2) / 2
5.2.4 Static Pressure Measurement
Accuracy*: Mean error is less than +/-5 mmHg Standard
Deviation is less than +/-8 mmHg
Range:
ADULT MODE PEDIATRIC MODE
40 to 255 mmHg 40 to 200 mmHg
Accuracy*: Mean error is less than ±5 mmHg, Standard
deviation is less than ±8 mmHg
Range:
ADULT MODE PEDIATRIC MODE
10 to 210 mmHg 10 to 150 mmHg
Range: 0 – 325 mmHg
Static Accuracy: ±3 mmHg over the entire range.