Patient Parameter Specifications Appendix
5 - 6 0070-10-0603-01 Duo™ Operating Instructions
5.2.6.6 NIBP Start Pressure Settings and Ranges
The Start Pressure is adjustable and is set to the following defaults:
5.2.6.7 NIBP Measurement Cycle
There is one mode of measurement operation: manual. The manual mode requires the
operator to initiate the measurement cycle.
In the manual mode, the unit adjusts the inflation pressure according to the previous systolic
pressure. After the first successful measurement is made, the subsequent inflation pressure
becomes +50 ±10 mmHg above the previous systolic pressure measurement.
5.2.7 SpO
2
Performance Requirements
The Duo monitor is capable of providing SpO
2
functional saturation level measurements via
an OEM Masimo MS-7 pulse oximeter, or an OEM Nellcor MP-506 pulse oximeter.
5.2.7.1 SpO
2
Agency Requirements
The SpO
2
function performs in accordance with the requirements of EN 865: 1997.
5.2.7.2 Masimo SpO
2
Performance Requirements
The Masimo MS-7 pulse-oximeter with SET technology is implemented.
SPO
2
PATIENT SIZE
PRESSURE INCREMENT
(DEPENDING ON ALGORITHM) DEFAULT START PRESSURE
Adult Mode 30 – 60 mmHg 178 ±5 mmHg
Pediatric Mode 30 – 60 mmHg 133 ±5 mmHg
Sensor Compatibility: Compatible with LNOP Series sensors.
Display Range: 1 – 100%
Resolution: 1%