Care and Cleaning of the Monitor User Maintenance
3 - 2 0070-10-0603-01 Duo™ Operating Instructions
3.2 Care and Cleaning of the Monitor
The monitor housing may be cleaned with a mild soap and water solution or ammoniated
window cleaner. Apply cleaning solution to the cloth, not directly onto the monitor. DO NOT
apply large amounts of liquid. DO NOT use abrasive cleaning agents or organic solvents.
WARNING: Do not clean the monitor while it is ON and/or connected to
AC power.
To prevent scratches on the screen, carefully remove dust and dirt particles with a fine, soft-
hair brush or a soft sponge moistened with cleaning solution. Fingerprints and stains may be
removed by using a liquid lens cleaner and a soft cloth. DO NOT wipe a dry screen or use
alcohol or a solvent containing chlorinated hydrocarbon.
3.3 Care and Cleaning of Accessories
3.3.1 SpO
2
Sensors
NOTE: Refer to the individual instruction sheets that are packaged
with each sensor.
1. Inspect the sensors and cables for damage on a daily basis. Replace as necessary.
2. Clean reusable sensors before and after each use as follows:
Wipe the patient contact area using a soft cloth with a mild soap and water solution, or
isopropyl alcohol. Hydrogen peroxide can be used to remove dried blood from all
accessible surfaces.
Clean the cable with a 3% hydrogen peroxide solution, isopropanol solution, or other
active reagent. Do not subject the connector of the sensor to such a solution.
Allow the sensor to completely dry before using.
CAUTION: When cleaning SpO2 sensors, do not use an excessive
amount of liquid. Wipe the sensor surface with a soft cloth,
dampened with a cleaning solution.
CAUTION: Do not subject the SpO2 sensor to autoclaving.
CAUTION: If the SpO2 sensor or patient cable are damaged in any
way, discontinue use immediately. To prevent damage, do
not soak or immerse the sensor in any liquid solution. Do
not attempt to sterilize.
CAUTION: Do not use SpO2 sensors or cables that are damaged or
have deteriorated.