Patient Parameter Specifications Appendix
5 - 12 0070-10-0603-01 Duo™ Operating Instructions
Sinusoidal Vibration: Per IEC 60068-2-6
1 g or 0.07 mm, 57 – 62 Hz crossover frequency
10 to 500 Hz, 10 sweep cycles in each axis
Random Vibration: Per IEC 60068-2-34
0.02 g2/Hz
20 – 500 Hz
Low degree of reproducibility
9 minutes per axis
Drop: The monitor meets the requirements specified by
ECRI PB-296 892 section AIII 3.3 for Class 3
devices.
Impact: The monitor meet the requirements specified by ECRI
PB-296892, section AIII 3.2 for Class 3 devices.