C-8 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 10L24EA
Imaging Mode: Color+B+PW-Mode/Power+B+PW-Mode
Index label
MI
TIS
TIB
TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
0.95
1.25
1.96
2.76
Index component value
1.25
1.18
1.25
1.96
Acoustic
Parameters
p
r,α
at z
MI
(MPa)
2.28
P (mW)
143.11
148.54
124.59
P
1
(mW)
41.64
42.93
z
s
(cm)
1.00
z
b
(cm)
0.69
z
MI
(cm)
0.60
z
pii,α
(cm)
0.60
f
awf
(MHz)
5.76
6.86
7.05
6.80
Other
Information
prr (Hz)
4,321.00
srr (Hz)
3.00
n
pps
/
I
pa,α
at z
pii,α
(W/cm
2
)
217.09
I
spta,α
at z
pii,α
or z
sii,α
(mW/cm
2
)
590.81
I
spta
at z
pii
or z
sii
(mW/cm
2
)
769.74
p
r
at z
pii
(MPa)
2.53
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
8.3cm
8.3cm
8.3cm
8.3cm
B Focus Position
1.0cm
5.0cm
1.0cm
5.0cm
Color SG Position
1.0cm
5.0cm
1.0cm
5.0cm
PW SV Position
1.0cm
5.0cm
1.0cm
5.0cm
B Working Frequency
8
8
8
8
C Working Frequency
5.7
5.7
5.7
5.7
PW Working Frequency
5.7
6.6
5.7
5.7
B PRF
1238
749
1238
749
C PRF
943
1189
943
1189
PW PRF
4321
4356
4321
4356
PW SV
0.5mm
0.5mm
0.5mm
0.5mm