Acoustic Output Reporting Table (60601-2-37) C-9
C.2 35C20EA
Transducer Model: 35C20EA
Imaging Mode: M-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.84
0.27
1.11
0.60
Index component value
0.25
0.27
0.60
1.11
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
1.32
P (mW)
33.77
33.97
34.04
P
1
(mW)
21.27
21.40
z
s
(cm)
2.32
z
b
(cm)
4.24
z
MI
(cm)
4.63
z
pii
,α
(cm)
4.63
f
awf
(MHz)
2.46
2.45
2.43
2.45
Other
Informatio
n
prr (Hz)
1,999.0
0
srr (Hz)
/
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
122.12
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
203.20
I
spta
at z
pii
or z
sii
(mW/cm
2
)
458.73
p
r
at z
pii
(MPa)
1.95
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
16.6cm
16.6cm
16.6cm
16.6cm
Focus position
6.0cm
6.0cm
6.0cm
6.0cm
Working Frequency
H6.0
H6.0
H6.0
H6.0
PRF
1999
1999
1999
1999

Table of Contents