C-10 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 35C20EA
Imaging Mode: B-mode/Smart3D/ Tissue Harmonic Imaging
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.84
0.35
0.96
0.96
Index component value
0.35
0.35
0.96
0.35
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
1.32
P (mW)
86.00
86.00
86.00
P
1
(mW)
21.67
21.67
z
s
(cm)
/
z
b
(cm)
/
z
MI
(cm)
4.53
z
pii
,α
(cm)
4.53
f
awf
(MHz)
2.45
3.37
3.39
3.38
Other
Informatio
n
prr (Hz)
4,073.0
0
srr (Hz)
33.00
n
pps
1.33
I
pa,α
at z
pii
,α
(W/cm
2
)
154.65
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
9.50
I
spta
at z
pii
or z
sii
(mW/cm
2
)
20.93
p
r
at z
pii
(MPa)
1.84
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
16.6cm
6.5cm
6.5cm
6.5cm
Focus position
6.0cm
4.0cm
4.0cm
4.0cm
Working Frequency
H6.0
3.5
3.5
3.5
PRF
4073
8992
8992
8992