C-28 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 65C15EA
Imaging Mode: Color+B-Mode / Power+B-Mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
1.10
0.10
0.13
0.25
Index component value
0.10
0.10
0.13
0.13
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
2.49
P (mW)
10.00
12.00
14.00
P
1
(mW)
3.91
4.69
z
s
(cm)
/
z
b
(cm)
/
z
MI
(cm)
0.50
z
pii
,α
(cm)
0.50
f
awf
(MHz)
5.07
5.99
5.95
5.40
Other
Information
prr (Hz)
1,906.00
srr (Hz)
9.00
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
317.09
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
6.37
I
spta
at z
pii
or z
sii
(mW/cm
2
)
7.59
p
r
at z
pii
(MPa)
2.72
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
11.1cm
11.1cm
11.1cm
11.1cm
B Focus Position
1.0cm
2.5cm
2.5cm
2.5cm
Color Sampling Gate
Position
1.0cm
2.5cm
2.5cm
2.5cm
B Working Frequency
6.5
6.5
6.5
5
C Working Frequency
5
5
5
4.4
B PRF
4756
4756
4756
5001
Color PRF
1906
1906
1906
2086