EasyManua.ls Logo

Mindray DP-50 Vet - Page 277

Mindray DP-50 Vet
343 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Acoustic Output Reporting Table (60601-2-37) C-29
Transducer Model: 65C15EA
Imaging Mode: Color+B+PW-Mode/Power+B+PW-Mode
Index label
MI
TIS
TIB
TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
0.94
0.41
1.53
0.98
Index component value
0.41
0.31
0.56
1.53
Acoustic
Parameters
p
r,α
at z
MI
(MPa)
2.10
P (mW)
25.99
31.31
28.55
P
1
(mW)
17.18
22.78
z
s
(cm)
1.44
z
b
(cm)
0.88
z
MI
(cm)
1.08
z
pii,α
(cm)
1.08
f
awf
(MHz)
5.03
5.26
5.99
6.08
Other
Information
prr (Hz)
3,474.00
srr (Hz)
3.00
n
pps
/
I
pa,α
at z
pii,α
(W/cm
2
)
249.29
I
spta,α
at z
pii,α
or z
sii,α
(mW/cm
2
)
645.86
I
spta
at z
pii
or z
sii
(mW/cm
2
)
962.79
p
r
at z
pii
(MPa)
2.44
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
11.1cm
11.1cm
11.1cm
11.1cm
B Focus Position
1.0cm
5.0cm
1.0cm
1.0cm
Color SG Position
1.0cm
5.0cm
1.0cm
1.0cm
PW SV Position
1.0cm
5.0cm
1.0cm
1.0cm
B Working Frequency
6.5
5
6.5
6.5
C Working Frequency
5
4.4
5
5
PW Working Frequency
5
5
5
5
B PRF
1876
1860
1876
1876
C PRF
635
630
635
635
PW PRF
3474
3445
3474
3474
PW SV
0.5mm
0.5mm
0.5mm
0.5mm

Table of Contents