C-36 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 65EB10EA_T
Imaging Mode: Color+B+PW-Mode/Power+B+PW-Mode
Index label
MI
TIS
TIB
TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
0.30
0.12
0.24
0.33
Index component value
0.12
0.11
0.14
0.24
Acoustic
Parameters
p
r,α
at z
MI
(MPa)
0.68
P (mW)
9.95
12.01
13.38
P
1
(mW)
5.52
6.47
z
s
(cm)
0.96
z
b
(cm)
0.98
z
MI
(cm)
0.98
z
pii,α
(cm)
0.98
f
awf
(MHz)
4.94
4.92
4.92
5.00
Other
Information
prr (Hz)
739.00
srr (Hz)
4.00
n
pps
/
I
pa,α
at z
pii,α
(W/cm
2
)
14.27
I
spta,α
at z
pii,α
or z
sii,α
(mW/cm
2
)
0.13
I
spta
at z
pii
or z
sii
(mW/cm
2
)
0.18
p
r
at z
pii
(MPa)
0.80
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
8.3cm
8.3cm
8.3cm
8.3cm
B Focus Position
2.0cm
2.5cm
2.0cm
1.0cm
Color SG Position
2.0cm
2.5cm
2.0cm
1.0cm
PW SV Position
2.0cm
2.5cm
2.0cm
1.0cm
B Working Frequency
6.5
5
5
5
C Working Frequency
5
4
4
4
PW Working Frequency
4
5
5
5
B PRF
2243
1493
2201
1493
C PRF
739
1459
725
1459
PW PRF
3498
3453
3432
3453
PW SV
0.5mm
0.5mm
0.5mm
0.5mm