Acoustic Output Reporting Table (60601-2-37) C-37
C.6 65EB10EA_S
Transducer Model: 65EB10EA_S
Imaging Mode: M-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.79
0.17
0.37
0.25
Index component value
0.17
0.11
0.26
0.37
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
1.94
P (mW)
7.37
5.51
7.63
P
1
(mW)
7.37
5.51
z
s
(cm)
1.33
z
b
(cm)
1.78
z
MI
(cm)
1.75
z
pii
,α
(cm)
1.75
f
awf
(MHz)
6.04
4.70
4.64
4.31
Other
Informatio
n
prr (Hz)
1,000.0
0
srr (Hz)
/
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
157.39
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
36.57
I
spta
at z
pii
or z
sii
(mW/cm
2
)
76.23
p
r
at z
pii
(MPa)
2.79
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
12.9cm
12.9cm
12.9cm
12.9cm
Focus position
2.0cm
5.5cm
2.0cm
5.0cm
Working Frequency
6.5
H8.0
H9.0
H9.0
PRF
1000
2999
2999
2999

Table of Contents