C-38 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 65EB10EA_S
Imaging Mode: B-mode /Smart3D/Tissue Harmonic Imaging
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.83
0.04
0.04
0.04
Index component value
0.04
0.04
0.04
0.02
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
2.04
P (mW)
4.00
2.00
2.00
P
1
(mW)
1.79
0.89
z
s
(cm)
/
z
b
(cm)
/
z
MI
(cm)
1.65
z
pii
,α
(cm)
1.65
f
awf
(MHz)
6.04
5.14
5.04
5.04
Other
Informatio
n
prr (Hz)
5,096.0
0
srr (Hz)
41.00
n
pps
1.33
I
pa,α
at z
pii
,α
(W/cm
2
)
158.75
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
4.50
I
spta
at z
pii
or z
sii
(mW/cm
2
)
9.36
p
r
at z
pii
(MPa)
2.75
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
12.9cm
6.5cm
6.5cm
6.5cm
Focus position
2.0cm
0.5cm
0.5cm
0.5cm
Working Frequency
6.5
5
5
5
PRF
5096
8992
8992
8992

Table of Contents