Acoustic Output Reporting Table (60601-2-37) C-83
Transducer Model: 50L60EAV
Imaging Mode: B+PW-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
1.30
0.87
2.01
1.82
Index component value
0.74
0.87
0.78
2.01
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
2.46
P (mW)
140.29
131.01
139.54
P
1
(mW)
40.46
42.67
z
s
(cm)
2.56
z
b
(cm)
5.56
z
MI
(cm)
2.71
z
pii
,α
(cm)
2.71
f
awf
(MHz)
3.59
4.28
4.44
4.16
Other
Informatio
n
prr (Hz)
699.00
srr (Hz)
33.00
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
359.42
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
267.74
I
spta
at z
pii
or z
sii
(mW/cm
2
)
526.53
p
r
at z
pii
(MPa)
3.25
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
12.9cm
12.9cm
12.9cm
12.9cm
B Focus Position
2.5cm
10.0cm
6.0cm
10.0cm
SV Position
2.5cm
10.0cm
6.0cm
10.0cm
B Working Frequency
4
4
4
4
PW Working Frequency
3.5
3.5
3.5
3.5
B PRF
4199
1276
1276
1258
PW PRF
699
3595
3595
3595
PW SV
0.5mm
0.5mm
0.5mm
0.5mm

Table of Contents