C-84 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 50L60EAV
Imaging Mode: Color+B-Mode / Power+B-Mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.87
0.07
0.07
0.24
Index component value
0.07
0.07
0.07
0.07
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
1.65
P (mW)
24.30
24.20
23.40
P
1
(mW)
4.05
4.03
z
s
(cm)
/
z
b
(cm)
/
z
MI
(cm)
3.16
z
pii
,α
(cm)
3.16
f
awf
(MHz)
3.60
4.12
4.12
4.12
Other
Information
prr (Hz)
631.00
srr (Hz)
11.00
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
192.43
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
6.12
I
spta
at z
pii
or z
sii
(mW/cm
2
)
13.46
p
r
at z
pii
(MPa)
2.35
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
16.6cm
16.6cm
16.6cm
16.6cm
B Focus Position
3.0cm
10.0cm
10.0cm
10.0cm
Color Sampling Gate
Position
3.0cm
10.0cm
10.0cm
10.0cm
B Working Frequency
4
4
4
4
C Working Frequency
3.5
3.5
3.5
3.5
B PRF
2775
2775
2775
2775
Color PRF
631
631
631
631

Table of Contents