29 - 6
The following table lists approved Masimo SpO
2
cable cleaning and disinfecting agents:
29.6 Sterilization
Sterilization is not recommended for this monitor, related products, accessories, or supplies unless otherwise
indicated in the Instructions for Use that accompany the products, accessories or supplies.
29.7 Cleaning the Thermal Print Head
Dirty print head deteriorates printing quality. Check the printout to ensure the printing is legible and dark. Light
printing may indicate a dirty print head.
To clean the thermal print head, follow this procedure:
1. Take measures against the static electricity, such as the wrist strap.
2. Remove the recorder module from the module rack.
3. Open the recorder door and remove the recording paper.
4. Gently wipe the print head with cotton swabs dampened with ethanol to remove the dust and foreign par-
ticles.
5. Wipe off excess moisture with dry cotton swabs.
6. Allow the print head air dry.
7. Reload the recording paper and close the recorder door.
Clorox Healthcare® Hydrogen Peroxide
Cleaner Disinfectant Wipes
Wipes Clorox professional products company
Diversey Oxivir® TB
Wipes
Wipes Diversey Inc
PDI Super Sani-Cloth®
Germicidal Disposable Wipe
Wipes PDI Inc.
VIRAGUARD®
Hospital Surface Disinfectant Towelettle
Wipes VERIDIEN corporation
Virex® TB Liquid, spray Diversey Inc
Glutaraldehyde, 2% Liquid /
Ethanol, 70% Liquid /
Isopropanol, 70% Liquid /
Sodium hypochlorite bleach, 0.5% Liquid /
Hydrogen peroxide, 3% Liquid /
Rely+On™ Virkon®
High Level surface Disinfectant, 1%
Powder Antec International Ltd
1-Propanol, 50% Liquid /
Product Name Product Type Manufacturer
Isopropanol Liquid Isopropanol 70%
Product Name Product Type Manufacturer