D - 11
D.2.16 Technical Alarm Messages Related to Networked Monitoring
Note: XX refers to the department name, YY refers to the room number, and ZZ refers to the bed number.
D.2.17 Other System Technical Alarm Messages
PDF storage space is
nearly full
Prompt / Delete the files saved under the PDF file path to
release storage space. Otherwise you cannot save
new PDF files.
Error storing PDF file Prompt / The PDF file path settings on the printer server
and the PDFCreator are not consistent or the PDF
storage space is full. Check the PDF file path
settings for consistency, or delete the files saved
under the PDF file path to release storage space.
Change the print server
language to be consistent
with this monitor
Prompt / Verify that the language settings of the printer
server and the monitor are consistent, Otherwise
you cannot perform printing.
Print Server Disconnected Prompt / Check that the monitor is properly connected
with the printer server.
Alarm message Default priority Indication on
alarm reset
Cause and solution
Alarm message Default priority Indication on
alarm reset
Cause and solution
CMS/eGW Disconnected Low B The monitor is disconnected from the CMS.
Check the network connection.
View Bed XX YY-ZZ,
Network Disconnected.
Low A The network is interrupted when the monitor is
viewing the remote device. Check the network
connection.
Viewed by Bed XX YY-ZZ,
Network Disconnected.
Low A The network is interrupted when the monitor is
viewed by another remote device. Check the
network connection.
WLAN IP Address Conflict Low C Wireless network IP network conflicts. Check the
network settings.
LAN1 IP Address Conflict Low C Wired network LAN1 IP network conflicts. Check
the network settings.
Fail To Get WLAN IP
Address
Low C Unable to automatically obtain the wireless
network IP address. Check the network settings.
Fail To Get LAN1 IP
Address
Low C Unable to automatically obtain the wired
network LAN1 IP address. Check the network
settings.
Alarm message Default priority Indication on
alarm reset
Cause and solution
XX: Disconnected
(XX refers to the name of
the external device)
High B Corresponding external device is disconnected.
Check the connection between the monitor and
the external device.
Data Card Err High C The storage card fails or files are damaged.
Restart the monitor to format the storage card. If
the alarm persists, contact your service
personnel.
Loading Default Config
Failed
Low A The default configuration is not correctly loaded.
The monitor will restore to the factory default
configuration for the current patient category.

Table of Contents