C - 12
C.1.7 C.O. Default Settings
C.1.8 CO
2
Default Settings
C.1.8.1 General Settings
Display Format Sys/Dia (Mean)
Display Alarm Limit Off
Use PA-D as PAWP Off
Item Default Setting
Item Default Setting
TB Alarm switch On
High limit 39.0 °C
Low limit 36.0 °C
Priority Med
Alarm Outputs Off
Comp Const 0.542
Auto Start On
Auto TI On
Item Default Setting
EtCO2 Alarm switch On
High limit Adult, Pediatric: 50 mmHg
Neonate: 45 mmHg
Low limit Adult, Pediatric: 25mmHg
Neonate: 30mmHg
Priority Med
Alarm Outputs Off
FiCO2 Alarm switch On
High limit 4 mmHg
Priority Med
Alarm Outputs Off
Apnea Delay Adult, Pediatric: 20 s
Neonate: 15 s
RR Source Auto
Speed 6.25 mm/s
Scale 50 mmHg
Waveform Type Draw