EasyManua.ls Logo

Mindray ePM 10M - C Default Settings; ECG, Arrhythmia, ST and QT Default Settings

Mindray ePM 10M
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C - 1
C Default Settings
C.1 Parameters Default Settings
C.1.1 ECG, Arrhythmia, ST and QT Default Settings
C.1.1.1 ECG Default Settings
Item Default Setting
HR/PR Alarm switch On
High limit Adult: 120 bpm
Pediatric: 160 bpm
Neonate: 200 bpm
Low limit Adult: 50 bpm
Pediatric: 75 bpm
Neonate: 100 bpm
Priority Med
Alarm Outputs Off
Extreme Tachy Alarm switch On
High limit Adult: 160 bpm
Pediatric: 180 bpm
Neonate: 220 bpm
Priority High
Alarm Outputs Off
Extreme Brady Alarm switch On
Low limit Adult: 35 bpm
Pediatric: 50 bpm
Neonate: 60 bpm
Priority High
Alarm Outputs Off
Alarm Source Auto
ECG1 II
ECG2 (5-lead, 6-lead, 12-lead) V, Va, V1
Va (for 6-lead only) Va
Vb(for 6-lead only) Vb
ECG Gain ×1
Speed 25 mm/sec
Filter
OR: Surgery
CCU: Diagnostic
Other departments: Monitor
High Freq Cut-off (for 12-lead analysis) 35 Hz

Table of Contents

Related product manuals