C - 10
IBP-M Alarm switch On
High limit
â–  Art/Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure
Adult: 110 mmHg
P
ediatric: 90 mmHg
Neonate: 70 mmHg
â–  PA
Adult: 20 mmHg
P
ediatric, Neonate: 26 mmHg
â–  ICP/RAP/LAP/UVP/P3/P4 venous pressure
Adult: 10 mmHg
P
ediatric, Neonate: 4 mmHg
â–  CVP
Adult: 14 cmH
2
O
Pediatric, Neonate: 5 cmH
2
O
Low limit
â–  Art/Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure
Adult: 70 mmH
g
Pediatric: 50 mmHg
Neonate: 35 mmHg
â–  PA
Adult: 0 mmHg
Pediatric, Neonate: 12 mmHg
â–  ICP/RAP/LAP/UVP/P3/P4 venous pressure
Adult: 0 mmHg
Pediatric, Neonate: 0 mmHg
â–  CVP
Adult: 0 cmH
2
O
Pediatric, Neonate: 0 cmH
2
O
Priority Med
Alarm Outputs Off
Art-S Extreme Alarm switch Off
High limit Adult: 175 mmHg
Pediatric: 130 mmHg
Neonate: 95 mmHg
Low limit Adult: 75 mmHg
Pediatric: 60 mmHg
Neonate: 50 mmHg
Priority High
Alarm Outputs Off
Art-D Extreme Alarm switch Off
High limit Adult: 105 mmHg
Pediatric: 80 mmHg
Neonate: 65 mmHg
Low limit Adult: 35mmHg
Pediatric: 30 mmHg
Neonate: 15 mmHg
Priority High
Alarm Outputs Off
Item Default Setting