Parameter Menus Operations
2 - 54 0070-10-0666-01 Trio™ Operating Instructions
Temperature Setup Menu Selections
Temperature Alarm Limits
Alarms occurring during the process of temperature measurement include two (2) types: 
physiological alarms and technical alarms. Physiological alarms occur when the patient’s 
temperature values are equal to or exceed set alarm limits. Technical alarms are any 
temperature-related alarms, which are not physiological, such as functional failures.
ALM Allows the user to turn the TEMP alarm ON or OFF. Choose ON to 
enable the alarm; choose OFF to disable the alarm. If the alarm is set 
to OFF, the alarm OFF symbol   will display to the right of TEMP 
on the screen. The TEMP alarm is activated when the temperature 
value exceeds set high or low TEMP values.
ALM PRIORITY Enables priority selection for the TEMP alarm. Choices are 1, 2 and 
3. Priority 1 alarms are considered the most serious.
ALM PRINT Enables or disables automatic printing during a TEMP alarm 
condition. Choose ON to enable printing upon TEMP alarm. Choose 
OFF to disable printing upon TEMP alarm.
TEMP ALM HI Allows the user to set the upper limit of Temperature 1 alarm. (See 
“Temperature Alarm Limits” on page 2-54.)
TEMP ALM LO Allows the user to set the lower limit of Temperature 1 alarm. (See 
“Temperature Alarm Limits” on page 2-54.)
TEMP UNIT Allows the user to select the unit of measure for temperature. The 
selections are degrees Celsius (°C) or degrees Fahrenheit (°F). 
RESTORE 
DEFAULTS
Allows the user to restore the TEMP user default configuration.
FIGURE 2-41 Confirmation Dialog Box
HIGH 
ALARM (ºC)
LOW 
ALARM (ºC)
HIGH 
ALARM (ºF)
LOW 
ALARM (ºF)
Temperature (T1) 35 – 43 [40] 26 – 38 [30] 95 – 109.4 [104] 78.8 – 100.4 [86]
* Factory default values shown in brackets.