Trio™ Operating Instructions 0070-10-0666-01 2 - 55
Operations Parameter Menus
2.4.5.2 Temperature Troubleshooting
MESSAGE/
PROBLEM REASON SOLUTION
T1 SENSOR OFF Temperature cable of channel 1 may
be disconnected from the monitor
Make sure that the temperature
cable is properly connected, check
connection between probe and
cable
TEMP SENSOR
OFF
Temperature cable of channel 1 may
be disconnected from the monitor
Make sure that the temperature
cable is properly connected, check
connection between probe and
cable
T1 ALM LMT ERR Functional failure Notify hospital technician or
Customer Support
TEMP ALM LMT ERR Functional failure Notify hospital technician or
Customer Support
T1 EXCEED Temperature value of channel exceeds
measurement range
Check patient, notify physician
TEMP EXCEED Temperature value of channel exceeds
measurement range
Check patient, notify physician
Temperature Probes
not working
Poor contact from probes to body Check the body surface contact at
the probe tip
Reposition or apply
thermoconductive gel
Temperature not
displayed
Cable not plugged in Check cable and probe

Table of Contents

Other manuals for Mindray Trio

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Trio and is the answer not in the manual?

Mindray Trio Specifications

General IconGeneral
BrandMindray
ModelTrio
CategoryMedical Equipment
LanguageEnglish

Related product manuals