Parameter Menus Operations
2 - 62 0070-10-0666-01 Trio™ Operating Instructions
2.4.6.3 IBP Troubleshooting
MESSAGE/
PROBLEM REASON SOLUTION
SENSOR OFF, FAIL Transducer may be disconnected
during attempt to zero
Check transducer and cables for
connection. If problem persists,
notify hospital technician or
Customer Support
IN DEMO FAIL Monitor is in DEMO mode during
attempt to zero
Turn monitor OFF. Turn ON to
return to normal monitoring mode
PRESSURE OVER
RANGE, FAIL
Stopcock not open to atmosphere Turn stopcock to atmosphere prior
to zero attempt. If problem persists,
notify hospital technician or
Customer Support
PULSATILE
PRESSURE, FAIL
Stopcock is not open to atmosphere
and a pulsatile pressure has been
detected
Turn stopcock to atmosphere prior
to zero attempt. If problem persists,
notify hospital technician or
Customer Support
IBP SENSOR OFF IBP cable may be disconnected from
monitor
Make sure that cable is properly
connected
IBP INIT ERR1 Module failure Notify hospital technician or
Customer Support
IBP COMM STOP Module failure or communication
failure
Notify hospital technician or
Customer Support
IBP COMM ERR Communication error Notify hospital technician or
Customer Support
IBP ALM LMT ERR Functional failure Notify hospital technician or
Customer Support
IBP SYS EXCEED Systolic pressure value exceeds
measurement range
Check patient, notify physician
IBP DIA EXCEED Diastolic pressure value exceeds
measurement range
Check patient, notify physician
IBP MAP EXCEED Mean arterial pressure value exceeds
measurement range
Check patient, notify physician
IBP NEED ZERO-
CAL
Zero calibration must be performed
before measuring IBP
Zero calibration must be
performed. Notify hospital
technician or Customer Support.
Dampened Invasive
Waveform
Air bubbles in tubing Eliminate air from tubing
Kinked invasive catheter Change position of catheter, check
patient
Catheter against wall of blood vessel Check for leaks at connector, flush
catheter
Blood in tubing Pump pressure bag up to
300 mmHg
Catheter partially occluded Consult physician
IBP not Displayed /
No Waveform
Cable not plugged in Check cable connections
Transducer not connected Check transducer connections
Stopcock turned improperly Check transducer for proper
alignment
Transducer not zeroed Check and zero the transducer
Abnormally high or
low readings
Transducer too HIGH or too LOW in
relation to patient's heart
Check patient, adjust transducer,
re-zero
To Purchase, Visit Avobus.com or call 1-800-674-3655

Table of Contents

Other manuals for Mindray Trio

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Trio and is the answer not in the manual?

Mindray Trio Specifications

General IconGeneral
BrandMindray
ModelTrio
CategoryMedical Equipment
LanguageEnglish

Related product manuals