Trio™ Operating Instructions 0070-10-0666-01 2 - 61
Operations Parameter Menus
IBP Alarm Limits
Alarms occurring during the process of IBP measurement include two (2) types: physiological 
alarms and technical alarms. Physiological alarms occur when the patient’s pressure values 
are equal to or exceed set alarm limits. Technical alarms are any IBP-related alarms, which 
are not physiological, such as functional failures.
IBP 
LABEL
SYS 
HIGH 
(mmHg)
SYS 
LOW 
(mmHg)
MEAN 
HIGH 
(mmHg)
MEAN 
LOW 
(mmHg)
DIA 
HIGH 
(mmHg)
DIA 
LOW 
(mmHg)
ART
Adult 
5 – 300 
[180]
0 – 150 
[80]
5 – 150 
[100]
- 5 – 100 
[40]
5 – 140 
[100]
0 – 120 
[50]
Ped 
5 – 240 
[150]
0 – 130 
[70]
5 – 100 
[80]
- 5 – 50 
[30]
5 – 100 
[80]
0 – 95
[40]
PA
Adult 
5 – 300 
[180]
0 – 150 
[80]
5 – 150 
[100]
- 5 – 100 
[40]
5 – 140
[100]
0 – 120
[50]
Ped 
5 – 240 
[150]
0 – 130 
[70]
5 – 100 
[80]
- 5 – 50
[30]
5 – 100
[80]
0 – 98
[40]
CVP
Adult 
n/a n/a 5 – 150
[10]
- 5 – 100
[0]
n/a n/a
Ped 
n/a n/a 5 – 100
[10]
- 5 – 50
[0]
n/a n/a
RAP
Adult 
n/a n/a 5 – 150
[10]
- 5 – 100
[0]
n/a n/a
Ped 
n/a n/a 5 – 100
[10]
- 5 – 50
[0]
n/a n/a
LAP
Adult 
n/a n/a 5 – 150
[10]
- 5 – 100
[0]
n/a n/a
Ped 
n
/a n/a 5 – 100
[10]
- 5 – 50
[0]
n/a n/a
ICP
Adult 
n/a n/a 5 – 150
[10]
- 5 – 100
[0]
n/a n/a
Ped 
n/a n/a 5 – 100
[10]
- 5 – 50
[0]
n/a n/a
* Factory default values shown in brackets.