Alarm Type Display Speaker Flash LED, Vibrator and Top Red LED
Major Fault
(Asic, EEPROM,
Detection, Data
Integrity) (*)
4 beeps
every second
FLASH LED 1 ash every second
VIBRATOR 1 vibration every second
TOP LED 3 red ashes every second
Alarms
5.6.2 Major Faults
The DMC 3000 conducts periodic self tests every 10 minutes in Run, Pause and Sleep modes. Most of the tests
are also performed every second in Run mode. All the main functions are controlled by the microcontroller and
all the data, parameters and congurations integrity are protected by the CRC16.
The calibrations and algorithm parameters are protected by the CRC16 and saved inside a specic and dedica-
ted EEPROM. The dierent DC voltages on the electronic board are periodically measured and controlled.
If a fault is detected, an alarm is generated and a message is alternately shown with the measure on the display.
A major fault prevents the DMC 3000 from going into RUN mode
(*) See 5.6.3 for the list of faults and solutions to perform