Manual Calibration & Maintenance
8.2 Maintenance
Mirion Technologies recommends to perform a complete check at least every year, including:
Calibration check
Display, top and ashing LEDs inspection
Check on wake from sleep, the dosimeter will perform a self test in which the following can be observed: the
speaker will chirp once, the front ash LED and the three (3) top LEDs (red, green, blue) will briey illuminate,
all display segments will light and the back light will turn on
Speaker check (turn on your DMC with rate warning set point at "0")
Note:
The DMC 3000 is provided with an initial factory calibration traceable to international reference standards.
Mirion recommends an annual calibration check; although the re-calibration interval may be determined by the customer based
on service environment, frequency of use and requirements dened by local regulations.