-22-
151153EN-G
DMC 3000 User's Guide
Alarms
5.1 Dose and Rate Alarms or Warnings
On DMC 3000 rmware V.7, the DMC 3000 has 2 congurable dose alarms from 1 µSv to 10Sv and 2 con-
gurable rate alarms from 10 µSv/h (*) to 10 Sv/h for Hp(10) (X + Gamma) deep dose, X + Hp(0,07) Gamma +
Beta shallow dose or Hp(10) Neutron dose and 2 congurable total dose alarms Gamma + Neutron.
(*) Please contact Mirion Technologies for a rate set point between 1 µSv/h and 10 µSv/h.
Dose and rate warnings are acknowledgeable by pressing either push button for longer than 3 seconds.
The rate alarm and warning occur when the rate exceeds the set point. The rate alarm and warning disap-
pear when the rate decreases to 80% of the set point if latched feature is enabled (see §2.3.5 and §2.3.6).
On DMC 3000 rmware V.7, the displayed value of dose or rate blinks simultaneously with the indicator
when it has generated the alarm or warning, see example below.
Note:
When displaying a measure (here, Total dose), the value and DOSE
ALARM are blinking, notifying an alarm on the total
Gamma+Neutron dose.