-8-
151153EN-G
DMC 3000 User's Guide
General
2.3 Functionalities
2.3.1 Dose, Rate Notification
The high contrast and backlit LCD display provides clear indication of dose, rate and other information with 8
digits and symbols. Whatever the unit there is no scale change for the whole range.
The data information displayed on the DMC 3000 is selectable using DMCUser software.
When in pause mode (no measurement) the DMC 3000 can display PAUSE, SLEEP, ASSIGN, the wearer name,
the calibration due date or a user message
Notication:
d: for dose
R: for rate
h: Run time (hh:mm:ss)
t: Remaining Time before alarm
V: Life dose or life time
M: maintenance dose or time
xxx dEF Default, "xxx" identies
the type of default (see §5.6.3
Troubleshooting guide)
Origin of dose/Rate displayed
(measurements or alarm thresholds).
: X+Gamma Hp(10)
: Neutron Hp(10)
: total dose ((X+Gamma) + neutron Hp(10))
: beta and X+Gamma Hp (0.07)
: Gamma rate in counts per second (PRD)
Blinking Alarm Indicator:
DOSE ALARM: Dose Alarm
ALARM RATE: Rate Alarm
DOSE : Dose pre-alarm
RATE: Rate pre-alarm
: Time Alarm
Dose, Rate unit can be:
µSv (/h),
mSv (/h),
mrem (/h)

Related product manuals