EasyManua.ls Logo

Mirion Technologies DMC 3000 - Compatibility with Mirion Technologies Products

Mirion Technologies DMC 3000
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
-6-
151153EN-G
DMC 3000 User's Guide
2.2 Compatibility with Mirion Technologies Products
The DMC 3000 features two protocols, G3 and G2, that use the same communication medium, with an omni-
directional, low-frequency (125 kHz) transceiver.
The enhanced "G3" communication protocol to support additional features and attached measurement
modules. It also includes a compatibility mode for previous Mirion Technologies products including Access
Control, Turnstiles and Telemetry devices.
The compatible DMC 2000 communication protocol "G2" is indicated on the DMC 3000 by an asterisk in
the display's bottom right corner. This guarantees compatibility with the existing Mirion readers (LDM 220,
LDM 2000 and LDM 3000) and software (DosiServ, DosiCare, DosiFFR and Sentinel).
Note: G2 protocol is available only on V5 rmware version.
The enhanced communication protocol "G3" requires new readers such as the LDM 320D, LDM 320W,
LDM3200 or LDM 2000 with upgraded rmware.
Attached measurement modules (neutron Hp(10), beta Hp(0,07), PRD...) can be added when the DMC 3000 is
in “G3” protocol and with rmware version V.7. See §10 for Add-on modules information.
Telemetry module can be added to the DMC 3000 whatever the protocol “G2” or G3”.
General
"G2" DMC 2000 compatible protocol

Related product manuals