-40-
151153EN-G
DMC 3000 User's Guide
Memory:
Data storage in EEPROM for more than 10 years
Historical record of dose increments and events
RTC for date stamping information
3800 history events at programmable intervals of 10 seconds, 1 minute, 10 minutes, 1 hour or 24 hours
in G2 protocol
Up to 10 days of record with 1 minute interval in G3 protocol with average ambiant rate of 5 µSv/h.
See table below:
9.7 Environment
In accordance with European regulation (Directive DEEE 2002-96-CE), dosimeters should not be thrown out
with non-sorted waste, but must be selectively collected.
The batteries should not be recharged or thrown in re.
Technical Characteristics
Total duration of record without overlap
(DMC 3000 in G3 protocol (rmware V7), with/without Neutron or Beta Hp(0,07) Module) (**)
Interval of dose stamp
Minimum Value : one increment of dose equal
or greater than 1µSv each interval in Primary and
secondary (if module) doses
Average Value: one increment of dose equal or
greater than 1µSv each 10 interval in Primary and
secondary (if module) doses
Average rate Without Module With Module (*) Average rate Without Module With Module (*)
10 seconds 360 µSv/h 4 hours 2 hours 32 µSv/h 42 hours 19 hours
1 minute 60 µSv/h 23 hours 10 hours 6 µSv/h 10 days 5 days
10 minutes 6 µSv/h 10 days 4 days 55 nSv/h 100 days 50 days
1 hour 1 µSv/h 60 days 26 days 10 nSv/h 640 days 290 days
24 hours 40 nSv/h 1400 days 630 days 4 nSv/h 15 000 days 7 000 days
(*) with Hp(0,07) Beta Module (worst case)