(7) The retained values are different from the values that were set.
Point to check Cause and remedy
Is there an error message about the
memory backup battery?
The data processing unit “QM-Data” is in a state in which it cannot
back up memory. Those part programs or measured data not
saved to a personal computer or floppy disk will be lost.
(8) Errors related to the Digital Scale unit occur frequently.
Point to check Cause and remedy
Is the ambient noise appropriate?
• Connect the AC adapter to a different power supply, one that does
not have large current flowing through it.
• Use the data processing unit “QM-Data” away from equipment
generating electromagnetic noise, for example a welding machine or
electric discharge machine.
• If the problem persists, it may be necessary to adjust the Digital Scale
unit. In that case, contact your supplier.
(9) It is not possible to print from the printer.
1) Confirm that the connector attached to the connection cable is securely connected.
2) Check whether or not an error message is displayed on the printer. If an error
message is displayed, correct the error by following the instructions in the printer
manual.
3) Disconnect the connection cable between the printer and the data processing unit,
then try the test print described in the printer manual.
No. 99MCA081A