e-Vision and e-Vision+ LP101013.100 – 01 January 2004
MKS Instruments, Spectra Products
21
Care should be taken in routing and securing all cables. Avoid running any of the
signal cables next to mains power cables or sources of electrical noise.
5.3 IP Address
The default IP address of the e-Vision is 192.168.0.250.
Details on how to change this IP address can be found in the manual supplied with
your RGA operating software.
5.4 Re-setting the IP address
To re-set the unit back to its default IP address, carry out the following procedure:
1. Push the “reset” button
2. Immediately push and hold the “mode” button until the “status” indicator
lights and flashes rapidly four times and is then extinguished.
3. Press the “reset” button to reboot the unit using the default IP address.