Item Description
Reading Modes Fluorescence
Transmitted Light (Brightfield)
Colorimetric
Objectives 4x
10x
20x (optional)
40x (optional)
63x (optional)
Fluorescence Color Channels FITC
DAPI (optional)
TRITC (optional)
CY5 (optional)
Texas Red (optional)
CFP (optional)
Environmental Control CO
2
:1% to 15% (optional)
O
2
:1% to 15% (optional)
Humidity:85% (optional)
Temperature Control
without environmental control cassette
6°C (11°F) above ambient up to 40°C (104°F)
with environmental control cassette
8°C (14°F) above ambient up to 40°C (104°F)
Ambient Operating Conditions Temperature (general use without environmental control): 18°C
to 30°C (64°F to 86°F)
Temperature (when using temperature control):18°C to 28°C
(64°F to 82°F); ambient temperature must be at least 8°C
(14°F)below the target temperature
Relative Humidity: 20% to 75% (non-condensing)
Ambient Storage Conditions Temperature: -20°C to +60°C (-4°F to 140°F)
Relative Humidity: 15% to 75% (non-condensing)
Altitude Restrictions Up to 2000m (6,562ft)
IEC Installation Category II
Pollution Degree 2
Data Connection Two (2) Ethernet ports