(3) Fresh Air Setup prompt (FAS?)/Pulse Check start screen (PULSE) is
displayed, depending on the sensor installed.
Fig. 4 Startup, continued
WARNING
Only carry out the Fresh Air Setup/bump test in fresh, uncontaminated air;
otherwise, inaccurate readings can falsely indicate a hazardous atmosphere
as safe. If the quality of the surrounding air is uncertain, do not carry out a
Fresh Air Setup/bump test. Do not use the Fresh Air Setup/bump test as a
substitute for daily calibration checks. The calibration check is required to
verify span accuracy. Failure to follow this warning can result in serious
personal injury or death.
XCell Pulse
Technology?
Is FAS
Enabled?
Startup
FASeuD laC
Pulse Check
Y
N
N
Y
Pulse Check
Start Screen
PULSE
Sensor Warmup
Screen
WAIT