ALTAIR 2X
20
Use
GB
FAS
FAS? is displayed.
To start the FAS press the button while FAS? is displayed.
The display will show the results of the setup, PASS or ERR.
Bump Test for Devices with XCell Pulse Technology
This test will start automatically.
- PULSE is displayed shortly.
- The electronic pulse check icon and a countdown are displayed.
- At the end of the countdown, the results of the electronic pulse check are
displayed (PASS or ERR).
If ERR:
Press the button to return to normal operating mode and perform a calibra-
tion.
If PASS:
- The display shows the Flow check icon and WAIT for followed by a count-
down.
At the end of the countdown, the display will show FLOW and the green LEDs
near the sensor will flash.
Exhale into the device.
Hold device close to mouth (5 cm/2 inches distance) and blow forcefully
onto the sensor opening for 2 seconds.
The results of the test are then displayed (PASS or ERR).
The fresh air setup has limits.
If a hazardous level of gas is present, the device displays ERR.
Press the button to acknowledge the error and perform a calibration.