ALTAIR 2X
21
Use
GB
If ERR:
(1) Press the button to return to normal operating mode and perform inspec-
tion of sensor.
(2) Repeat the bump test if necessary.
(3) If the bump test fails, calibrate the device ( Section 3.13).
If PASS:
- The device returns to normal operating mode.
- A is displayed for 24 hours to indicate that the self-test was successful.
- If the safe LED (green) option is enabled, the green LEDs will flash periodi-
cally.
To clean the sensor inlet before or after a flow test, only use approved cleaner
described in section 7.
(4) To bypass FAS, do
not press the button.
The device continues the ON sequence.