Response times with calibration cap and a flow rate of 1l / min
Combustible
Response time Recovery time
t50
*)
t90
*)
t50
*)
t10
*)
1-Ethoxy-2-Pro-
panol
20s 48s 20s 48s
Hydrogen 10s 18s 10s 18s
Methane 10s 18s 10s 18s
Propane 12 20s 12 20s
The response time will be increased up to 60 seconds by using the Sen-
sor gard.