The pressure in the test gas tube increases with rising test gas flow rates. A flow
rate of 1500 ml/min results in a pressure up to 600 hPa. This has to be considered
if a flow meter is used which is affected by the media pressure.
8.4 Flow Through Adapter
Fig. 16 Flow through adapter
The flow through adapter is for use with a pumped sampling system.
Material St. St. 316
Dimensions (in mm) 60 x 122 (diameter x height)
Weight 0.6 kg
Working temperature -30 °C to +70 °C
Storage temperature -30 °C to +70 °C
Max. wind speed up to 6 m/s
Applicable test gases H
2
, CH
4
, C
3
H
8
, CO and H
2
S in air, O
2
in N
2
(approval of other gases by MSA on request)
Recommended flow rate 1.0 l / min
Minimum flow rate 0,8 l / min
Maximum flow rate 1.5 l / min
Response times
CH
4
t
50
15 s t
90
40 s
C
3
H
8
t
50
20 s t
90
55 s
O
2
t
50
10 s t
90
45 s
CO t
50
15 s t
90
45 s
H
2
St
50
20 s t
90
60 s
Gas inlet/outlet thread: 1/8” NPT