EasyManua.ls Logo

MSA ULTIMA X5000 - Page 42

MSA ULTIMA X5000
95 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2. SelectInstrument.
3. ScrollandselectMin/Max/Average.
4. SelectSetInterval.
5. EnterInterval(1h,8h,24h)andSave.
6. ScrolltoSetStartHour.
7. EnterStartHour(0-24h)andSave.
TheMin/Max/AveragevaluesareonlyviewablethroughtheX/SconnectapporHART.
SwapDelay
SwapDelayallowstheuserabriefwindowtochangeanXCellsensorwithoutthedevicegoingintoafaultcondition.Oncea
sensorisdisconnectedfromthetransmitter,theuserwillhave2minutestoreconnectasensor.Duringthistime,thedevice
analogoutputwillgotoitsMaintenancelevel.Ifasensorisreconnectedorreplacedduringthe2minutewindow,thenew
sensor’scountdownsequencewillbeginandtheanalogoutputwillremainattheMaintenancelevel.Afterthesensor
countdowniscomplete,theanalogoutputwillreturntoreportingalivegasreading.Ifasensorisnotreconnectedafterthe
2minutewindow,theULTIMAX5000willentera“SensorMissing”faultcondition.AllXCellSensorshaveSafeSwapanddo
notneedtobedisconnectedfrompowerwhilechangingsensors.Formoredetailsonhowtochangesensors,see6.2
ReplacinganXCellSensor.SwapDelayisenabledonallULTIMAX5000transmittersbydefault.
NOTICE
Thetransitiontomaintenancemodeduringthe2minuteSwapDelaywindowandsensorcountdownwillnottriggerthe
FaultRelay.TheFaultRelaywillonlybetriggeredwhenthedeviceentersafaultcondition.
ToenableordisableSwapDelay:
1. ScrolltoSettings.
2. SelectInstrument.
3. ScrollandselectSwapDelay.
4. SelectEnabledorDisabled.
5. SelectSave.
TimeandDateSetup
TimeanddatearesetatthefactoryinGMT.Whenselected,thecurrentdateisdisplayed.SelectChangetoeditdateand
time.Theusermustsavetomoveontothenextdatesetting.TheMin/Max/Averagesettingsaredrivenbythedateandtime
andshouldbechangedtolocaltimefordataaccuracy.
TochangeTimeandDate:
1. ScrolltoSettings.
2. SelectInstrument.
3. ScrollandselectSetDate.
4. ScrollandSelectChange.
5. SelectYearandSave.
6. SelectMonthandSave.
7. SelectDayandSave.
US Ultima X5000 Gas Monitor 42
4 Operation

Other manuals for MSA ULTIMA X5000

Related product manuals